Sterility Testing: A Comprehensive Guide

What is Sterility Testing?

Sterility testing is a microbiological examination method used to determine whether a product that is required to be sterile has been contaminated by microorganisms. It is commonly applied in the pharmaceutical, biotechnology, and medical device manufacturing industries.

The Purpose of Sterility Testing

Sterility testing is applicable to most pharmaceutical products mandated to be sterile, including injectable drugs, eye drops, solid pharmaceuticals, and various sterile medical devices. Its scope of application is broad. If these products are contaminated with microorganisms, they may reduce efficacy or even cause infections and other severe complications upon contact with the human body. Therefore, the primary goal of sterility testing is to ensure that products designated as sterile, after undergoing aseptic manufacturing or terminal sterilization processes, are free of viable microorganisms, ensuring safety and effectiveness in clinical use.

Definition of Sterility and International Standard

Definition of Sterility

Sterility refers to the absence of viable microorganisms in a sample, which is a fundamental requirement for ensuring the safety of pharmaceutical products. The internationally recognized benchmark for sterilization efficacy is the Sterility Assurance Level (SAL). A product is considered sterile if SAL ≤ 10⁻⁶, meaning the probability of microbial contamination is less than or equal to one in a million.

International Standards for Sterility Testing

Standards such as ISO 11737-2, USP <71>, EP, JP, and the Chinese Pharmacopeia define the requirements for sterility testing, including test environments, sample preparation, and operating procedures. These ensure compliance with global standards and provide reliability for test results.

How is Sterility Testing Performed?

Two commonly used methods for sterility testing are the Membrane Filtration Method and the Direct Inoculation Method. According to the Pharmaceutical Inspection Co-operation Scheme (PIC/S), both methods must be performed under aseptic conditions in a Grade A cleanroom within a Grade B environment. The choice of method depends on the sample characteristics.

 

Method 1: Membrane Filtration Method

This is one of the most widely used sterility testing methods, suitable for liquid samples, solid samples, and medical devices with pipelines that can be rinsed and filtered.

Principle: The sample is filtered through a membrane with a pore size of less than 0.45 µm to capture potential microorganisms. The membrane is then transferred to an appropriate culture medium for incubation to detect the presence of viable microorganisms.

Equipment for Membrane Filtration Method

Equipment for Membrane Filtration Method

The required equipment includes vacuum pumps, filtration manifolds, and membrane filters. Reusable stainless-steel filter cups can be autoclaved or flame sterilized with a Lab burner to save sterilization time.

Highlight Products: Multi-Branch Filtration System Series!

MultiVac 301 – MB – A Multi-Branch Filtration System

MultiVac 301 MB A N scaled 1

- Save time & space
- Quick-assembly and disassembly
- Innovative vent valve design
- Offers 3-port and 6-port options

SolarVac 1201 – MB – T Vacuum Filtration System

SolarVac 1201 MB T N scaled 1

- Innovative rotary design
- Individual control valve
- Durable and autoclavable
- Offers 6-port and 12-port options

Operating Steps for Membrane Filtration Method

The operational process of the membrane filtration method for sterility testing can generally be divided into four stages: sample pre-treatment, sample filtration, membrane incubation, and result interpretation. The following will provide an explanation of each stage in sequence.

1.Sample Pre-treatment: Depending on sample type, pre-treatment is required before filtration to improve flow rate or prepare the sample for filtration.

Sample Type
Filterable Liquids
Ointments and Oils
Solid Antibiotics
Medical Devices
Pre-Treatment Method
Dilute with sterile liquid if too viscous.
Dissolve in isopropyl myristate, heat if necessary (up to 44°C).
Dissolve in sterile solution.
Rinse with sterile solution and filter the rinsate.

2.Sample Filtration: Pour the pre-treated sample into the filter funnels. Use a vacuum pump to create negative pressure, enabling vacuum filtration. For larger quantities, multi-branch filtration manifolds can be used.

3.Membrane Incubation: Transfer the membrane to a culture medium using sterile tweezers, seal, and incubate at an appropriate temperature for 14+ days.

4.Result Interpretation: Observe the culture medium for microbial growth. If no changes occur, the sample is sterile and complies with sterility testing requirements. If the medium becomes turbid, it indicates the presence of viable microorganisms, meaning the sample does not meet sterility requirements.

Method 2: Direct Inoculation Method

Direct Inoculation Method is designed for samples that cannot be processed using membrane filtration, such as insoluble solids or medical devices that are unsuitable for filtration.

The principle involves directly placing the sample into a culture medium and monitoring for microbial growth to assess sterility.

Operating Steps for Direct Inoculation Method

The operational process of the direct inoculation method can generally be divided into three stages: Sample preparation, culture medium incubation, and result interpretation. The following provides an explanation of each stage in sequence.

 

1.Sample Preparation:After preparing the sample, place it into the culture medium.

Sample Type
Solid Sample
Liquid Sample
Medical Devices
Pre-Treatment Method
Use an appropriate solvent to prepare a suspension, ensuring the sample can evenly contact the culture medium.
Directly add an appropriate amount to the culture medium for incubation.
Immerse the entire device or a suitable part of it directly in the culture medium, ensuring all surfaces that may come into contact with microorganisms are fully covered by the culture medium.

2.Culture medium incubation: Seal the culture medium and incubate for 14+ days.

3.Result Interpretation: Observe the culture medium for turbidity to determine sterility. If no changes occur, the sample is sterile and complies with sterility testing requirements. If the medium becomes turbid, it indicates the presence of viable microorganisms, meaning the sample does not meet sterility requirements.

Comparison of Methods: Membrane Filtration and Direct Inoculation

The choice between the Membrane Filtration and Direct Inoculation methods depends on sample characteristics to ensure accuracy and reliability.

Sterility Testing
Membrane Filtration Method
Direct Inoculation Method
Applicable Samples
Liquid samples, soluble samples, medical devices with tubing
Samples unsuitable for membrane filtration, such as high-viscosity solutions or specialized medical devices
Operating Principle
Uses a membrane with pore size ≤0.45 μm to capture microorganisms, followed by membrane incubation
Directly cultures the sample in a medium and observes for microbial growth
Required Equipment
Membranes, vacuum pump, manifold filtration system, stainless steel filter holder, sterilization equipment (e.g., flame sterilization with a bunsen burner or autoclaving)
Culture medium, sterile containers
Incubation Days
14 days or more
14 days or more

Summary of Sterility Testing

Sterility testing is an essential microbiological examination in pharmaceuticals and biotechnology. Using the Membrane Filtration and Direct Inoculation methods, aseptic packaged products can be effectively tested for contamination, ensuring the safety and stability of drugs and medical devices.

This article was originally published by our partners Rocker Scientific. View the original article here